Digital Oscilloscope DS-8000 Series
Various Functions

Smoothing function

Smoothing function

Upper section: Signal where noise is superimposed
Interruption: Smoothing ± 2 points
Lower section: Smoothing ± 6 points
Simple moving average processing allows the main signal to be measured by highlighting it versus the signal with waveform distortion. From the sampling data, the average value is calculated and displayed by shifting the average value for each certain section (which can be arbitrarily set by width).


Arithmetic function

Arithmetic function

Example: Response characteristics of inductor
Upper section: Voltage change (CH1)
Middle section: Current change (CH3)
Lower section: Power change=MATH(CH1*CH3)
Using the arithmetic function, loss, impedance, etc. can be determined. In addition, as with channel input (voltage/current), MATH has a rescale function, so you can directly read the characteristic values of the device.


FFT function

FFT function

The frequency components of the input signal can be determined.
In the example in the left figure, the time axis waveform has a 500 ns period signal superimposed on the square wave signal of the 100 ns period in the upper section. In the lower section, a 10 MHz odd-order harmonic 1-3-5 order is clearly shown by FFT analysis, and a 2 MHz signal can be seen between the frequency components.


Filter function

 Input signal

① Analog filter20MHz

② Digital filter(LPT)200MHz

③ Digital filter(HPF)35MHz

For each channel, the analog filter and the digital filter can be set separately.

① Analog filter function (when 20 MHz is set)
Input amplifier is band-limited to eliminate high-frequency noise.

②③ Digital filter function (when ② LPF200MHz, ③ HPF35MHz is set)
The filter type and cutoff frequency can be set, and the filter function is realized by digital signal processing.
In the example, one signal can be extracted from a signal in which two frequency components are superimposed.


Average operation

Noise superimposed waveform

Equalized waveform

The left upper waveform is a superimposed waveform of noise. In the upper right waveform, the noise is reduced by averaging. Capture the trigger as a single "Then, after capturing the data for the set number of averages, stop the measurement.


High Resolution mode, Advanced High Resolution(Advanced High Resolution)

Noise superimposed waveform

Equalized waveform

High Resolution mode
When set to a sample rate lower than the highest sampling, the data captured in the highest sampling is averaged and displayed in high resolution. You can effectively increase the vertical resolution by attenuating random noise. It can also be used for single-shot signals and repeating signals, and supports up to 16 bits.

Advanced High Resolution
Digital processing provides higher resolution and less noise than High Resolution.


sin (x)/X interpolation

In this mode, by creating interpolation data by curve fitting between acquired sampling points using sin (x)/x interpolation, the apparent sampling speed is increased and measured. Unlike equivalent sampling, it can also be used for single-shot signals.


Peak detection mode

Detects and displays the maximum and minimum values that occur within the interval that is twice the set sampling period, so that peak values such as noise waveforms can be reliably understood.

The above figure shows the capture method at the time of normal sampling (upper section) and at the time of peak detection (lower section).

If peak detection is not set, the waveform data point A in the figure may not be detected. Because it captures at 400 ps cycles independent of the sampling period, it can reliably capture phenomena that occur within the sampling period.


Roll sample

The waveform data is displayed in real time, and continuous waveform data is automatically scrolled from the left side of the screen to the right side. When using the maximum memory length of 120M points, very long level fluctuations of 5-50s/div can be seen continuously. The figure is an example that visually captures frequent amplitude fluctuations.
Roll operation is possible from 100 ms/div to 50 s/div with a memory length of 1.5k points as the fastest setting.



Cursor functions


The cursor functions are Time, Amplitude, Time & Amplitude, and Value at cursor. You can move the cursor individually or track two cursors simultaneously. The figure is an example of pulse measurement with the Time & Amplitude cursor. You can view the values in the CURSOR area of the cursor menu, horizontal axis, and channel menu. Note that the cursor window of the measurement screen can also be hidden so that it does not overlap the waveform.


Split waveform display


The vertical axis (amplitude) of each channel can be displayed independently. While maintaining the full scale resolution of the vertical axis, the Y-T display (vertical axis: amplitude horizontal axis: time) is displayed for up to 16 sections in the 8CH model (CH display 8 + MATH operation 8).
For example, while looking at the original waveform, you can observe integration, differentiation, smoothing, and FFT analysis at the same time.


Search function


Up to 30,000 waveforms can be searched in the section that is displayed on the screen. Since the position of the waveform abnormality is easy to identify, it also leads to a reduction in debugging man-hours. The figure shows the second pulse of 13.12 ns or less, which is detected and enlarged to the lower section of the screen with the waveforms before and after.


Measurement gate function


You can set each parameter operation in a specific section. As shown in the figure above, if you want to continuously check the change in pulse width at a specific timing while monitoring the waveform, you can display the respective pulse width by applying a measurement gate to each pulse. In addition, the change of the signal at different times in different channels can also be checked numerically.


Deskew function


Depending on the probe used, skewing may occur between channels. The Deskew function allows you to compensate for time differences between channels. This function is necessary for accurate loss analysis (voltage x current) and timing measurement. You can check numerically.


Sequence function/History function


The Sequence function is a high-speed trigger that allows you to set the horizontal axis sample mode to a sequence and store the data in the segmented memory, greatly reducing the data read time. The figure shows an example in which the 1/5-32,768th waveform is captured and the maximum value of each waveform is displayed.

The History function is always running in sample mode with no special settings in real-time mode. You can easily read the waveform history. The maximum number of records is 32,768 waveforms (dependent on memory). Historical data is displayed by specifying the index number of the history.


Data storage and offline analysis

By saving the measurement conditions and measurement data of the DS-8000, and later using the DS-8000 Viewer to read out the data, you can play back the waveform under different conditions, convert the waveform data to the required format (binary→ CSV format) and save it, and save the screen display in png format. To transfer long memory data to a PC, you can save via a USB3.0 storage device, or use the Ethernet or USB interface between the PC and the DS-8000.


The figure is an example of measuring current and voltage drop in the reactor.
The data is saved and the waveform is read out on the PC. Since the knobs etc. on the oscilloscope are displayed visually on the PC, you can perform analysis with the feeling of actually operating the oscilloscope.


External trigger function (factory option)

Compatible with external trigger input by selecting the AUX OUT external trigger input and the option (factory option applied at the time of shipment). You can capture the waveform using the timing of the control signal from the outside, such as system embedding.

Parameter for measurement

Measure type Measure content Units
Vertical(Vertical axis direction)
Maximum Maximum value of the waveform V
Minimum Minimum value of the waveform
Peak-Peak Difference between maximum and minimum values
Top Upper flat portion of the waveform
Base Lower flat portion of the waveform
Top-Base Difference between Top and Base (amplitude)
RMS RMS value of waveform
Cycle RMS RMS value of waveform per cycle
Mean Mean value of waveform
Cycle Mean Mean value of waveform per cycle
+OverShoot Overshooting in waveform rise %
-OverShoot Overshooting in waveform fall
Horizontal (horizontal direction)
Transition Time Waveform rise/fall time s
Tr 20-80% 20-80% rise time of waveform
Tf 80-20% 80-20% fall time of waveform
Tr10-90% 10-90% rise time of waveform
Tf 90-10% 10-90% fall time of waveform
Frequency Waveform frequency Hz
Period Waveform period s
+Pulse Count Number of positive pulses -
-Pulse Count Number of negative pulses -
+Pulse Width Width of positive pulse s
-Pulse Width Width of negative pulse s
Duty Cycle Waveform duty cycle (positive pulse width/period) %
Other
dV/dt Slope of the rising/falling edge of the waveform V/s
Integral Integral of the waveform Vs
Integral (Absolute) Absolute value of integral of the waveform
Integral (Positive) Integral with only positive values of the waveform (negative values are treated as 0)
Integral(Negative) Integral with only negative values of the waveform (positive values are treated as 0)
Skew (%) Time difference between edges of waveform
(Specify the threshold for edge detection in %)
s
Skew Level Time difference between edges of waveform
(Specify the threshold for edge detection in real numbers)
Phase (%) Phase difference of the waveform (specify the threshold for edge detection in %) Degree,
Radian,
%
Phase Level Phase difference of the waveform (specify the threshold for edge detection in real numbers)

Example of ideal probe for DS-8000

Voltage Probe

Order number Bandwidth MHz Attenuation Maximum voltageVpeak DS-8000 Series terminator Ω DS-8000 Series power supply connection Probe power supply
External power supply Battery-driven
BumbleBee
(Differential)
400 500/250/100/50:1 ±2,000 50 × Not supported PS-02 〇 AP-01
PS-03
SS-320 (differential) 100 500/50:1 ±1,400 1M ○Yes ○PS-25 × Not supported
PHV2000 400 100;1 4,000 1M
PHVS2000 400 1000;1 4,000 1M
TETRIS 1500 1,500 10:1 ±8V 50 × Not supported ○TETRIS
Standard Accessory
× Not supported
PML-711i-RO 500 10:1 300(CATII) 1M

Current probe

Order number Bandwidth
(MHz)
Sensitivity(mV/A) Maximum current
(Apeak)
DS-8000
Series
Terminator (Ω)
DS-8000
Series
Power supply connection
Probe power supply
External power supply Battery-driven
Rogowski Coil Current Probes
SS-280A
SS-280A-H
30 200~0.2 30~30,000 1M × Not supported
Option
AC adapter

AA
Batteries
4 pcs
SS-290 10/20 5~0.05 1,200~120,000 1M
SS-620 30 100~0.2 60~30,000 1M
SS-660
SS-660P
30 100~5 60~120,000 1M
SS-680 100 10~0.1 120~12,000 50
Clamp current probe (hole element type)
SS-531 120 0.1 / 1 / 10 30 / 5 / 0.5 1M
Option
Conversion cable
SS-200IW

Option
PS-54
×
Not supported
SS-521 120 1 5
Option
PS-52
PS-54
SS-550 100 0.1 30
SS-560 10 0.01 150
SS-570 2 0.01 500
Current transformer CT Series
58,59 Series

~30

0.01~1 ~500 50
13 Series ~60 0.05~1 ~100 50

Skew Calibrator

To accurately measure the instantaneous switching loss, it is necessary to align the phase (skew) between the voltage probe and the current probe as close as possible to the actual measured voltage and current amplitude.
The skew calibrator IE-1360 is a signal generator that outputs both voltage and current pulses at the same time. It can output up to 250V and 10A so that it can calibrate with sufficient amplitude even for high voltage and high current probes used for power electronics measurement. Note that skew adjustment uses the Deskew function on the oscilloscope.