Digital Multimeters VOAC7523H/7522H/7520H/7521H Performance
* Accuracy: ±(X% of reading + Y digits) is written as X + Y. Accuracy for all is 1 year after calibration.
1. Typical sample rate and resolution
Sample rate | Resolution | Measurement count | Hum rejection |
---|---|---|---|
SLOW | 5.5-digit | Approx. 4 times/second | ○ |
MID | 5.5-digit | Approx. 20 times/second | ○ |
FAST | 4.5-digit | Approx. 100 times/second |
×
|
2. DC voltage measurement (DCV)
The 50mV range is VOAC 7523H/7522H only.
Range | Resolution | Input resistance | Accuracy * | ||
---|---|---|---|---|---|
5.5-digit | 4.5-digit | SLOW/MID | FAST | ||
50mV | 0.1μV | 1μV | 100MΩ or more | 0.025+10 | 0.025+15 |
500mV | 1μV | 10μV | 1000MΩ or more | 0.012+5 | 0.012+10 |
5V | 10μV | 100μV | 0.012+2 | 0.012+7 | |
50V | 100μV | 1mV | Approx. 10MΩ | 0.016+5 | 0.016+10 |
500V | 1mV | 10mV | 0.016+2 | 0.016+7 | |
1000V | 10mV | 100mV |
The accuracy of the 50mV and 500mV ranges after zero correction by the REL operation
Sample rate of 50mV range is SLOW/MID: approx. 0.5 times/second, FAST: approx. 50 times/second
Maximum allowable voltage: 50mV to 5V ranges ±800V continuous
: 50V to 1000V range ±1100V continuous
Resolution and noise rejection
Resolution | Sample rate | NMRR | CMRR |
---|---|---|---|
5.5-digit | SLOW | 55dB or more | 120dB or more |
5.5-digit | MID | 55dB or more | 120dB or more |
4.5-digit | FAST | 0dB | 55dB or more |
3. CH-B DC voltage measurement (DCV)
VOAC7523H/7520H only
Range | Resolution | Input resistance | Accuracy * | |
---|---|---|---|---|
4.5-digit | SLOW/MID | FAST | ||
5V | 100μV | CH-B:H~CH-B:L 10MΩ±3% CH-B:H~CH-A:L 5MΩ±3% CH-B:L~CH-A:L 5MΩ±3% |
0.025+2 | 0.025+30 |
50V | 1mV | 0.025+8 | ||
300V | 10mV | 0.025+5 |
Maximum allowable voltage: ±300V, ±300V between CH-A L and CH-B
Resolution and noise rejection
Resolution | Sample rate | NMRR | CMRR | Influence of voltage between CH-A and L |
---|---|---|---|---|
4.5-digit | SLOW/MID | 55dB or more | 120dB or more | 56dB or more |
4.5-digit | FAST | 0dB | 55dB or more |
4. AC voltage measurement (ACV, DC + ACV) True RMS value detection
* VOAC 7521H/7520H up to 100kHz
Range | Resolution | Measurement range | Input resistance | |
---|---|---|---|---|
5.5-digit | SLOW | MID/FAST | ||
500mV | 1μV | 15Hz~300kHz * | 200Hz~300kHz * | Approx. 1MΩ// 100pF or less |
5V | 10μV | |||
50V | 100μV | |||
500V | 1mV | 45Hz~100kHz | 200Hz~100kHz | |
750V | 1mV | 45Hz~20kHz | 200Hz~20kHz |
Accuracy: SLOW/MID/FAST sample
Frequency | Accuracy * |
---|---|
15Hz~45Hz | 0.5+150 |
45Hz~100Hz | 0.25+150 |
100Hz~30kHz | 0.2+150 |
30kHz~100kHz | 0.5+300 |
100kHz~300kHz | 2.5+1000 |
For sine waves in the range of 5% to 100%
Coefficient for inputs other than sine waves
Frequency | Crest factor | ||
---|---|---|---|
1~1.5 | 1.5~2 | 2~3 | |
15Hz~30kHz | 0.05% | 0.15% | 0.30% |
30kHz~300kHz | 0.20% | - | - |
Response time
Sample rate | Resolution | Measurement count | Response time |
---|---|---|---|
SLOW | 5.5-digit | 4 times/second | 3 seconds or less |
MID/FAST | 5.5-digit | 20 times/second | 2 seconds or less |
Maximum allowable voltage: 780Vrms, ±1100VDC (continuous)
For DC + ACV, add 500 (45Hz or less) or 300 (45Hz or more) to the accuracy digits.
The sample rate FAST is the same as MID approx. 20 times/second.
5. DC current measurement (DCA)
Range | Resolution | Accuracy * | Input resistance | ||
---|---|---|---|---|---|
5.5-digit | 4.5-digit | SLOW/MID | FAST | ||
5mA | 10nA | 100nA | 0.05+7 | 0.05+17 | 150Ω or less |
50mA | 100nA | 1μA | 15Ω or less | ||
500mA | 1μA | 10μA | 2Ω or less | ||
10A | 100μA | 1mA | 0.2+7 | 0.2+17 | 0.1Ω or less |
The input terminals of the 5mA to 500mA range and the 10A range are different, so the automatic range between them is not performed.
Maximum allowable current: 5mA to 500mA range500mA (FUSE 0.5A/250V)
: 10A range10A (FUSE 15A/250V)
6. AC current measurement(ACA, DC+ACA)
Range | Resolution | Measurement range | Input resistance | |
---|---|---|---|---|
5.5-digit | SLOW/MID | FAST | ||
5mA | 10nA | 15Hz~5kHz | 200Hz~5kHz | 150Ω or less |
50mA | 100nA | 15Ω or less | ||
500mA | 1μA | 45Hz~5kHz | 2Ω or less | |
10A | 100μA | 0.1Ω or less |
Accuracy: SLOW sample
Frequency | Accuracy * | Crest factor | ||
---|---|---|---|---|
1~1.5 | 1.5~2 | 2~3 | ||
15Hz~45Hz | 1+200 | 0.05% | 0.15% | 0.30% |
45Hz~1kHz | 0.4+200 | |||
1kHz~5kHz | 5.0+200 |
5% of the range (10% of the 10A range) to 100% for sine waves
Response time
Sample rate | Resolution | Measurement count | Response time |
---|---|---|---|
SLOW | 5.5-digit | 4 times/second | 3 seconds or less |
MID/FAST | 5.5-digit | 20 times/second | 2 seconds or less |
Maximum allowable current: 5mA to 500mA range500mA (FUSE 0.5A)
: 10A range10A (FUSE 15A)
Less than or equal to the maximum allowable current including the DC component that is superimposed on the input current.
For 45Hz to 1kHz in the 10A range, add 0.3 to the % item.
For DC + ACA, add 500 (45Hz or less) or 300 (45Hz or more) to the accuracy digits.
The sample rate FAST is the same as MID approx. 20 times/second.
7. Resistance measurement(2WireΩ/4WireΩ)
4WireΩ is VOAC 7522H/7521H only
Range | Resolution | Accuracy * | Measured current | ||
---|---|---|---|---|---|
SLOW/MID | FAST | SLOW/MID | FAST | ||
50Ω | 0.1mΩ | 1mΩ | 0.025+10 | 0.025+15 | Approx. 10mA |
500Ω | 1mΩ | 10mΩ | 0.014+3 | 0.014+8 | Approx. 10mA |
5kΩ | 10mΩ | 0.1Ω | Approx. 1mA | ||
50kΩ | 0.1Ω | 1Ω | Approx. 100μA | ||
500kΩ | 1Ω | 10Ω | 0.015+3 | 0.015+33 | Approx. 10μA |
5MΩ | 10Ω | 10Ω | 0.033+30 | 0.033+30 | Approx. 1μA |
50MΩ | 100Ω | 100Ω | 0.25+30 | 0.25+30 | Approx. 100nA |
500MΩ | 1kΩ | 1kΩ | 1.5+50 | 1.5+50 | Approx. 10nA |
Maximum allowable voltage: ± 500Vpeak
Terminal open voltage: 12V or less
The range of 50Ω to 5kΩ is the accuracy after zero correction by the REL operation.
FAST in the 5MΩ to 500MΩ range is the same as MID, approx. 20 times/second.
8. Lo-Power resistance measurement (2WireΩ)
Range | Resolution | Accuracy * | Measured current | |
---|---|---|---|---|
SLOW/MID/FAST | SLOW/MID | FAST | ||
500Ω | 10mΩ | 0.1+5 | 0.1+15 | Approx. 1mA |
5kΩ | 0.1Ω | Approx. 100μA | ||
50kΩ | 1Ω | Approx. 10μA | ||
500kΩ | 10Ω | 0.2+30 | 0.2+40 | Approx. 1μA |
5MΩ | 100Ω | 0.2+30 | Approx. 100nA | |
50MΩ | 1kΩ | 1.5+30 | 1.5+30 | Approx. 10nA |
Maximum allowable voltage: ± 500Vpeak
Terminal open voltage: 12V or less
The range of 500Ω to 5kΩ is the accuracy after zero correction by the REL operation.
FAST in the 5MΩ to 50MΩ range is the same as MID, approx. 20 times/second.
SLOW/MID/FAST are all 4.5 digits.
9. Diode measurement
Measured current | Measurement range | Accuracy * | Terminal open voltage | Maximum allowable voltage |
---|---|---|---|---|
Approx. 1mA or 10mA | 0.1mV~5.0999V | 0.014+13 | 12V or less | ±500Vpeak |
10. Temperature measurement
Thermocouple | Measured temperature range | Accuracy * | Resolution | Maximum allowable voltage |
---|---|---|---|---|
R | -50~0℃ | 0.2+70 | 0.1℃ | ±500Vpeak |
0~+100℃ | 0.2+50 | |||
+100~+1768℃ | 0.2+30 | |||
K(CA) | -200~-100℃ | 0.15+50 | ||
-100~0℃ | 0.15+35 | |||
0~+1372℃ | 0.15+20 | |||
T(CC) | -200~-100℃ | 0.15+50 | ||
-100~0℃ | 0.15+35 | |||
0~+400℃ | 0.15+20 | |||
J(IC) | -200~-100℃ | 0.15+50 | ||
-100~0℃ | 0.15+35 | |||
0~+1200℃ | 0.15+20 | |||
E(CRC) | -200~-100℃ | 0.15+50 | ||
-100~0℃ | 0.15+35 | |||
0~+1000℃ | 0.15+20 |
The above accuracy does not include the accuracy of the thermocouple.
The operating environment temperature of the device is ±0.1°C/°C at 0°C to 18°C and 28°C to 50°C (total thermocouple).
The measured temperature -200°C or less may be displayed, but the accuracy is not guaranteed.
The reference thermoelectric power is calculated according to the broken line approximation according to JIS C 1602-1995.
When the device and the thermocouple are connected through a plug, the error of the contact temperature compensation is added.
After the thermocouple is connected, heat run for at least 1 hour. (Do not expose the device to the wind)
11. Frequency measurement
Sample rate | Measurement count (gate time) | Number of digits to display and measurement range | Accuracy * | |
---|---|---|---|---|
SLOW | Approx. 0.5 times/second (1s) |
6-digit | 15.0000Hz~1.00000MHz | 0.02+2 |
MID | Approx. 4 times/second (100ms) | 5-digit | 15.000Hz~1.0000MHz | |
FAST | Approx. 10 times/second (10ms) | 4-digit | 150.00Hz~1.000MHz |
AC coupling, Crest factor: 3 or less
Input attenuator uses ACV AUTO range.
Maximum allowable voltage:780Vrms, ±1100Vpeak
12. Combination of dual functions
*1 In the table, the number on the right shoulder of the△ mark indicates the corresponding “Constraint” number.
*2 CH-B DCV is only installed on VOAC7520H/7523H.
*3 4WΩ is only installed on VOAC7521H/7522H.
* Dual-function operation has the following limitations compared to single-function operation.
- You cannot select a sampling rate.
- The sampling rate will be delayed. (For example, in the case of a dual function with a combination of DCV and DCA, the limit is about 1 time/second. Please contact us for more details.)
- The display does not show the results of comparator operations, the type of statistical data, or attribute information such as functions, ranges, and timestamps.
- The main unit of measurement is not displayed. The units of measurement of the sub-side are displayed with the LED of “SUB DISP UNITS.”
13. General performance
Operations | Moving average, scaling operations, decibel operations (dBm, dBμ), differential operations, statistical operations (MAX, MIN, x, σ), comparison operations, dual function arithmetic operations, continuity tests | |
---|---|---|
Memory | DATA | 3000pcs data with 10msec resolution timestamp (elapsed time) |
SET UP | 10 pieces | |
Interface (Full remote) |
Standard | RS-232 |
Option | LAN, GP-IB | |
Power supply | Voltage | 100VAC (standard) 110V, 220V, 240V (option) |
Frequency | 50Hz, 60Hz | |
Power consumption | 21VA or less (including option) | |
Operating temperature and humidity | 0°C to 50°C (80% RH or less) However, there shall be no condensation and the between 40°C and 50°C, humidity shall be 70% RH or less. |
|
Storage temperature and humidity | -20°C to 60°C (70% RH or less) However, there shall be no condensation. It also includes the operating temperature range. |
|
Size | Shape | 210Wx99Hx353L mm (Optionally built into the main unit) |
Weight | 3.5kg or less (including option) | |
Accessories | Current measurement spare fuse, test leads (1 set), adjustment screwdriver, simple manual, instruction manual (CD-ROM), AC power cable |